Description
Technical description
Displacement | |
50Hz | 9.7 m3h-1 / 5.7 ft3min-1 |
60Hz | 11.7 m3h-1 / 6.9ft3min-1 |
Speed (Pneurop 6602) | |
50Hz | 8.5 m3h-1 / 5 ft3min-1 |
60Hz | 10 m3h-1 / 5.9 ft3min-1 |
Ultimate pressure | 2 x 10-3 mbar / 1.5 x 10-3 Torr |
Ultimate pressure GB I | 3 x 10-2 mbar / 2.3 x 10-2 Torr |
Ultimate pressure GB II | 6 x 10-2 mbar / 4.6 x 10-2 Torr |
Ultimate pressure with Fomblin® | 2 x 10-2 mbar / 1.5 x 10-2 Torr |
Max inlet pressure for water vapour | 38 mbar / 28.5 Torr |
Max water vapour pumping rate – GB I | 60 gh-1 |
Max water vapour pumping rate – GB II | 220 gh-1 |
Max allowed outlet pressure | 0.2 bar gauge / 2.8 psig |
Max allowed inlet and gas ballast pressure | 0.5 bar gauge / 7 psig |
Motor power 50/60Hz | 450 / 550W |
Power connector 1-ph | IEC EN60320 C13 |
Nominal rotation speed 50/60Hz | 1500 / 1800rpm |
Weight (without oil) | 28 kg / 61.6 lb |
Oil capacity min/max | 0.45 / 0.75 litres |
Recommended oil | Ultragrade 19 |
Inlet flange | NW25 |
Exhaust flange | NW25 |
Noise level | 48 dB(A) @ 50 Hz |
Operating temperature range | 12 – 40 °C |
Dimensions
Applications
- Analytical instruments
- Centrifuges, ultra-high speed
- Coating
- Degassing/curing – oil, epoxy resin
- Distillation/extraction/filtration
- Freeze drying
- Furnaces
- Gas bottle filling/emptying
- Gel drying
- Glove boxes, ovens
- Laboratory bench top vacuum
- Leak detectors, Helium
- Load locks and transfer chambers
- Refrigeration manufacture
- Research and development
- Rotary, centirfugal evaporators
- Scanning electron microscopes – SEM
- SEM/FIB (ion beam repair)
- Solvent recovery
- Surface science instruments
- Turbomolecuar backing pumps
Technology
All our RV pumps are quiet. Noise levels of 48 dBA 50Hz have been achieved by extensive work on the pump design, drive train and motor.
The pumps have the unique ability to deliver excellent ultimate vacuum in both high vacuum and high throughput modes with or without gas ballast. This gives the user an unrivalled opportunity to select the optimum pumping performance without compromising the application.
The simple to use mode selector and accessible, three-position gas ballast control allow the pump to be reconfigured at any time, even when the pump is running.
In high vacuum mode, the RV pump is ideal for analytical instruments, electron microscopes, physics research, backing turbo pumps and leak detection applications. Configure the same pump in the high throughput mode, and it becomes suitable for distillation, laboratory furnaces, backing vapor pumps, solvent concentration, freeze drying and other drying applications.
Features and Benefits
High reliability – generous lubrication of moving parts.
Safe process and systems – fast acting inlet valve for best in class anti‑suck back protection.
Configured to meet your vacuum needs – mode selector and three position gas ballast
No unplanned downtime – designed for easy maintenance.
Better work environment – quiet running and intrusive frequencies minimised.
Three position gas ballast allows either 60 g/hr or a class leading 220 g/hr (290 g/hr on RV12) of water vapour to be handled.
Universal motor to cover the widest possible voltage and frequency range from a single variant.
Smooth controlled inlet valve opening by oil pressure to eliminate pressure surges.
Rapid closing inlet valve within 0.4 seconds of pump stopping.